Chapter 1
Laboratory Preparation
Gas and Plumbing Requirements
You must provide the gas supplies for your gas chromatograph. Be sure to order
your gases and regulators far enough ahead of time to have them ready for the GC
installation process.
WARNING! All Thermo Fisher Scientific gas chromatographs are designed to use an inert gas
as the carrier gas. If you wish to use hydrogen as a carrier gas, you must install a
hydrogen sensor. Contact a Thermo Fisher Scientific sales representative if you
plan to use hydrogen as the carrier gas in your new FOCUS GC. If you don't have the
hydrogen sensor, you must use an inert carrier gas.
Table 1-7 lists the gas recommendations for capillary columns.
Table 1-8 lists the gas recommendations for packed columns.
32
Table 1-7. Capillary and Wide-Bore Column Gas Recommendations
Detector Type
Carrier Gases
FID
Helium
Nitrogen
Hydrogen
TCD
Helium
Nitrogen
Hydrogen
Argon
Table 1-8. Packed Column Gas Recommendations
Detector Type
Carrier Gases
TCD
Helium
Nitrogen
Hydrogen
Argon
Gas and Plumbing Requirements
Fuel Gases
Hydrogen + Air
Helium
Nitrogen
None
Same as carrier
Fuel Gases
None
Not Required
Site Preparation and Installation Manual
Make-Up Gas
Make-Up Gas