14 Maintenance

14 Maintenance

14.1 General instructions

14.2 Intervals

Part concerned
Device
Battery
Charging station
352
EN
MEDUCORE Standard
Risk of injury from impaired device functions or defects, lack
of testing and excessive service life!
If the device is not correctly tested, or is operated for too long,
impaired device functions, deviations or defects may occur. The
device cannot be used correctly, and may injure the patient and
the user.
 Observe the intervals for maintenance, for Technical Safety
Check ("Sicherheitstechnische Kontrolle" in accordance with
§ 11 of the German regulation MPBetreibV), and for
metrological check.
 Do not operate the device for longer than 10 years.
Maintenance, Technical Safety Check (in Germany only), and
measures such as inspections and repairs must only be carried out
by the manufacturer or by a technician expressly authorized by it.
Interval
Maintenance-free
Technical Safety Check every year Manufacturer or a technician
Metrological check every two years
Maintenance-free
When stored in the device: Charge every 3 months.
When not stored in the device: Charge batteries with serial numbers
< 20,000 every 5 months and batteries with serial numbers ≥ 20,000
every 9 months.
Recommendation: Replace battery after 2 years.
Maintenance-free
Recommendation: Replace
charging station after 12 years
2
Maintenance by
expressly authorized by
manufacturer