6.13 Using audio alarm output

6.13.1 Canceling audio alarm output
Requirement
An alarm is active and is audible.
1. Briefly (< 2 s) press the alarm button
Result
Audio alarm output is canceled for this alarm. The symbol
appears in the display and no audio signal is outputted for this
alarm.
6.13.2 Pausing/muting audio alarm output
1. Press and hold the alarm button
Result
Audio alarm output pauses for the time set in the operator menu
(Operator menu | Alarm settings | Pause audio). The symbol
to ∞ (infinite), audio alarm output is permanently paused (audio
alarm output is muted). The symbol
A reminder signal can remind you at certain intervals that audio
alarm output is paused or muted. You can set the reminder signal
in the operator menu (Operator menu | Alarm settings |
Reminder signal).
the NIBP cuff using the Stop function button.
appears in the display. If you set the time in the operator menu
(> 2 s).
appears in the display.