Weinmann MEDUCORE Standard2 Instructions For Use Manual page 319

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
SpO
Infant alarm limit
settings
NIBP systolic
NIBP systolic
NIBP diastolic
NIBP diastolic
Pause audio
-
Audible alarm
emitted in
Manual mode
(only with
-
Manual
defibrillation
option)
Possible
values
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
saturation
66 % - 100 %
2
saturation
65 % - 99 %
2
45 mmHg -
130 mmHg,
in increments of 5
40 mmHg -
125 mmHg,
in increments of 5
25 mmHg -
100 mmHg,
in increments of 5
20 mmHg -
95 mmHg,
in increments of 5
1 min
2 min
5 min
10 min
∞ (infinite)
Activated
Deactivated
13 Operator menu
Description
Here you can set the upper
( ) and lower ( ) limit
values for the patient group
as of which the device is to
emit an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value
above the set value for
the lower alarm limit.
The setting range for the
lower alarm limit ends at
least one setting value
below the set value for
the upper alarm limit.
Here you can set the length of
time for which audio alarm
output is paused. If you select
∞ (infinite), audio alarm
output is paused
permanently (audio alarm
output is muted).
Here you can set whether
audio alarm output is active
or inactive when the user
calls up manual mode.
MEDUCORE Standard
Factory
setting
200/min
100/min
200/min
100/min
95 %
85 %
100 mmHg
50 mmHg
70 mmHg
30 mmHg
2 min
Deactivated
2
EN
319