7 Disassembling
Result
7.3 Disassembling pulse oximetry sensor
Requirement
222
EN
MEDUCORE Standard
4. Dispose of used defibrillation electrodes.
5. When used in a protective transport bag: Guide the master
cable through the opening between the side compartment and
the device compartment of the protective transport bag.
The defibrillation electrodes/paddles and the master cable have
been disassembled.
and pulse oximetry sensor connecting
cable
•
The device is switched off
page
119).
•
The device is disconnected from the patient.
1. When used in a protective transport bag: Open the viewing
window and side compartments of the protective transport
bag.
2. Disconnect the pulse oximetry sensor connecting cable with
the selected pulse oximetry sensor from the SpO
2
(see "6.2 Switching the device off",
connection.
2