Printer function restricted due to excessive distance between
printer and device!
An excessive distance between the printer and the device may
lead to special characters in the printout and confuse the user.
 Reduce the distance between the printer and the device.
1. Position the printer within range of the device.
When doing so, please note: The printer must be positioned at
least 20 cm away from the patient's body (excluding limbs).
2. Press and hold the On/Off button
The status indicator of the printer is green and an audible signal
sounds.
3. Wait until the printer has connected to the device.
When doing so, please note:
When connection is successful, the symbol
the display.
The device outputs an audio signal when connected
successfully.
The connection indicator appears in the printer display
when connection is successful.
4. Leave the printer switched on.
When doing so, please note:
The printer battery gets hot during operation.
Maximum print time may be reduced at low temperatures.
If the device is switched off, the connection to the printer
is disconnected. The printer switches off automatically after
1 minute.
5. If the paper roll has finished: Insert new printer paper.
6. If the printer battery is empty: Insert a charged printer battery.
7. To disconnect the connection between the printer and the
device: Press and hold the On/Off button
for 3 s.
The printer switches off.
Result
The printer has been connected to the device.
of the printer for 3 s.
MEDUCORE Standard
6 Operation
appears in
of the printer
2
EN
197