6 Operation
128
EN
MEDUCORE Standard
6. Select the electrode surface area of the paddles to suit the
patient group:
Adults: Use attachments for large electrode surface areas.
Children/infants: Twist off and remove attachments for
large electrode surface areas.
The attachments for small electrode surface areas are
located under the attachments for large electrode surface
areas.
2
Electrode surface
area for adults
Electrode surface
area for children/
infants