6 Operation
6.13.3 Canceling muting or pausing of audio alarm
Requirement
Result
6.14 Changing the volume of the device
Requirement
Result
6.15 Saving event manually in the session
194
EN
MEDUCORE Standard
output
An alarm is active and is muted or paused.
1. Briefly (< 2 s) press the alarm button
Alternatively:
Switch mode.
Muting or pausing of audio alarm output is canceled.
•
The device is switched on
page
118).
•
A patient group is selected
page
120).
1. Open the application menu using the navigation knob.
2. Select the Volume menu item.
3. Select volume and confirm with the navigation knob.
The volume of the device has been changed.
data set
Measured values and user actions carried out on the device are
saved in the internal memory and on the SD card.
With the event button
the device automatically (e.g. intubation, administration of
medication, etc.) can be saved in the data set in order
retrospectively to assign them chronologically during evaluation.
2
.
(see "6.1 Switching on the device",
(see "6.4 Selecting patient group",
, events which cannot be recorded by