Weinmann MEDUCORE Standard2 Instructions For Use Manual page 258

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

10 User menu
Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
SpO
Alarm limit
settings Infant
SpO
SpO
NIBP systolic
NIBP systolic
Automatic alarm
-
limits
258
EN
MEDUCORE Standard
Possible
values
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
saturation
66 % - 100 %
2
saturation
65 % - 99 %
2
45 mmHg -
saturation
130 mmHg,
2
in increments of 5
40 mmHg -
saturation
125 mmHg,
2
in increments of 5
25 mmHg -
100 mmHg,
in increments of 5
20 mmHg -
95 mmHg,
in increments of 5
10 %
20 %
30 %
2
Description
Here you can set the upper
( ) and lower ( ) limit
values as of which the device
is to trigger an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value above
the set value for the lower
alarm limit. The setting range
for the lower alarm limit ends
at least one setting value
below the set value for the
upper alarm limit.
Here you can set the
automatic alarm limits. The
device automatically specifies
the alarm limits for the
physiological alarms. The
deviation is 10 %, 20 % or
30 % from the measured
values at the time of
activation.
Factory
setting
200/min
100/min
200/min
100/min
95 %
85 %
100 mmHg
50 mmHg
70 mmHg
30 mmHg
20 %