18 Technical data
18.13 Non-invasive blood pressure
Specification
Manufacturer
Measuring method
Measurement range (systolic):
•
Adults
•
Children
•
Infants
Measurement range (diastolic):
•
Adults
•
Children
•
Infants
Accuracy (during operation between 0 °C and 50 °C) ±3 mmHg
18.14 Operation/data management
Specification
Display
Audio output
Session documentation
Data transmission via SD card
Data transmission via WiFi
Data transmission via Bluetooth
Data evaluation
380
EN
MEDUCORE Standard
measurement (NIBP monitoring)
®
interface
2
Non-invasive blood pressure
measurement
SunTech Medical
Oscillometric, diastolic values correspond to phase 5
Korotkoff sounds
40 mmHg to 260 mmHg
40 mmHg to 230 mmHg
40 mmHg to 130 mmHg
20 mmHg to 200 mmHg
20 mmHg to 160 mmHg
20 mmHg to 100 mmHg
Operation/data management
•
Illuminated symbols
•
Device status indicators
•
Voice prompts
•
Alarm tones
•
Audible signals
Automatic recording of measured values for ECG,
SpO
and NIBP and of event data
2
SD card with 32 GB
•
Data transfer types: 12-lead ECG, function check
results, session data, service data
•
WiFi: 802.11a/b/g/n
•
Security: WPA2-PSK
•
Protocol: https
•
Port: 443
Data transfer type: Session data
Via DEFIview PC software