Weinmann MEDUCORE Standard2 Instructions For Use Manual page 256

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

10 User menu
Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
SpO
Alarm limit
settings Adult
NIBP systolic
NIBP systolic
NIBP diastolic
NIBP diastolic
256
EN
MEDUCORE Standard
Possible
values
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
saturation
66 % - 100 %
2
saturation
65 % - 99 %
2
45 mmHg -
260 mmHg,
in increments of 5
40 mmHg -
255 mmHg,
in increments of 5
25 mmHg -
200 mmHg,
in increments of 5
20 mmHg -
195 mmHg,
in increments of 5
2
Description
Here you can set the upper
( ) and lower ( ) limit
values as of which the device
is to trigger an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value above
the set value for the lower
alarm limit. The setting range
for the lower alarm limit ends
at least one setting value
below the set value for the
upper alarm limit.
Factory
setting
120/min
50/min
120/min
50/min
100 %
85 %
220 mmHg
75 mmHg
110 mmHg
35 mmHg

Advertisement

Table of Contents
loading

Table of Contents