6 Operation
136
EN
MEDUCORE Standard
Setting
Audio outputs
CPR voice prompts
Metronome
CV ratio
Metronome frequency
Start analysis automatically
Start analysis automatically Activated
After cardiac rhythm analysis and shock delivery (if necessary), the
device instructs you to carry out CPR. A metronome provides a
guide for chest compressions.
Voice prompt (optional)
Carry out cardiopulmonary
resuscitation
1. Carry out chest compressions:
30 chest compressions for the adult patient group
15 chest compressions for the child patient group
Continuous chest compression on intubated patients
When doing so, please note:
The metronome sets the ideal frequency.
The device only issues certain AED instructions and voice
prompts after CPR is complete (Operator menu | Adult
AED settings/Child AED settings | CPR phase |
Duration), as complete and correct carrying out of
cardiopulmonary resuscitation takes priority.
2
Patient group
Adult
Deactivated
15:2/30:2
30:2
100/min
AED instruction
CPR
Child
Deactivated
15:2/30:2
15:2
100/min
Activated