Mobile station for wm 100 td therapy devices (68 pages)
Summary of Contents for Weinmann MEDUMAT EasyCPR
Page 1
This document is not a substitute for the description of the function check in the instructions for use. For comprehensive information, please refer to the instructions for use of the device. The instructions for use are available on the Internet at www.weinmann-emergency.com. Perform the function check prior to every use.
Page 2
Warning Devices and accessories which are defective or not ready for use can disrupt the therapy or cause it to fail completely! Perform a function check prior to every use. Only use devices and accessories which have successfully passed the function check.
Contents Warning and overview of device 1. Preparing for the function check 2. Visually checking the device and components 3. Checking the system for leaks 4. Checking visual and audio alarm output 5. Checking the supply pressure alarm 6. Checking the ventilation rate 7.
Page 4
MEDUMAT Easy Overview of device Compressed gas connection Ventilation pressure display pMax button with LEDs for setting of max. ventilation pressure Connection for pressure measuring tube Connection for ventilation hose...
Page 5
Alarm field Alarm mute button with LED Battery compartment Manual button with LED for activation of manual mode Adjusting knob for ventilation settings On/Off button Connection for MEDUtrigger...
1. Preparing for the function check The device is connected to the oxygen supply. The ventilation hose, pressure measuring tube and MEDUtrigger are connected to the device. The testing bag is connected to the patient valve. 2. Visually checking the device and components The device and accessories do not show any signs of damage, deformations or soiling.
3. Checking the system for leaks OPEN CLOSE The needle on the gauge remains steady for one minute. 4. Checking visual and audio alarm output OPEN An audible signal is emitted when the device is switched on. Once the device is switched on, all the LEDs flash at least once and the lowest LED on the ventilation pressure display lights up green.
5. Checking the supply pressure alarm CLOSE After the oxygen cylinder has been closed, the visual and audible alarm for < 2.7 bar O is triggered once the device supply pressure falls below 2.7 bar O OPEN The < 2.7 bar O alarm stops when there is sufficient supply pressure. 6.
7. Checking the tidal volume and pAW alarm Select the following settings: Via the adjusting knob: Vt = 950 ml | Freq. = 10 bpm Via pMax button with control LEDs: pMax = 45 mbar Information: In combination with the testing bag, these settings can cause the pAW alarm to be triggered.
8. Checking the pAW /apnea alarm The device triggers the pAW /apnea alarm after the second mechanical breath. 9. Checking MEDUtrigger Select the following settings: Via the adjusting knob: Vt = 950 ml | Freq. = 10 bpm Via pMax button with control LEDs: pMax = 45 mbar The Manual button and the LEDs on MEDUtrigger light up.
Pressing the button on MEDUtrigger triggers a mechanical breath. 10. Checking the demand flow mode The demand flow LED lights up. After the testing bag has been pressed and released, a clacking sound is heard. The function check is passed successfully when all of the steps have been completed.