9. Maintenance

9.1 Intervals

Both filters must be regularly checked for
soiling.
The coarse filter should be cleaned once
a week and changed every 6 months.
The fine filter should be changed after
250 operating hours at the latest.
From reasons of hygiene we recommend:
replacing the pressure measurement
tube
after 6 months and earlier when
11
soiled;
replacing the complete mask system
every 6 to 12 months depending on
soiling;
replacing the exhalation system accord-
ing to the relevant operating instructions.
In addition, servicing must be carried out as a
preventative measure. Either after:
5000 operating hours (maintenance
symbol
appears in display);
26
2 years at the latest (see servicing sticker
on bottom of unit).
Maintenance includes the following:
Filter change;
Cleaning unit;
Replacement of any defective parts;
Complete check of unit functions and
pressure indicator.