Weinmann MEDUCORE Standard2 Instructions For Use Manual page 127

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

2. Connect the Pad connector of the paddles to the master cable.
When doing so, please note: The Pad connector must be
plugged in firmly.
Risk of injury due to air/moisture between the paddles and
the patient's skin!
Air (e.g. in the case of hirsute patients) or moisture between the
paddles and the patient's skin prevent correct shock delivery and
may result in burns to the skin and unsuccessful defibrillation/
cardioversion.
 Remove hair from hirsute patients.
 Rub the patient's skin dry.
 Wipe down oily skin with an alcohol pad.
 Always use electrode gel with paddles.
 Always press paddles on firmly with a contact pressure of about
8 kg.
3. Remove hair from the torso.
4. Rub damp spots on the torso dry.
5. Wipe down oily skin with an alcohol pad.
MEDUCORE Standard
6 Operation
2
EN
127

Advertisement

Table of Contents
loading

Table of Contents