Weinmann MEDUCORE Standard2 Instructions For Use Manual page 277

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

11.2.3 Low-priority alarm (turquoise)
Alarm
Battery operation
(Alarm deactivated after 10 s)
Check seating of SpO
Erroneous date
Insert battery
(Monitor mode)
Insert SD card
Internal memory full
Plug in SpO
sensor
2
Printer battery empty
Cause
Line power too low
Line power disconnected by
removing from the wall mounting
Power outage
Pulse oximetry sensor not
attached/not attached correctly to
sensor
2
the patient
Date read-out error
Internal module defective
Real Time Clock battery defective
or empty
Battery not inserted/not inserted
correctly
No SD card present/detected
No SD card inserted and session
duration > 4 h-10 h
SpO
connector of the pulse
2
oximetry sensor connecting cable
not connected/not connected
correctly to the device
Status of printer battery very low
11 Alarms and faults
Remedy
Restore line power.
Attach pulse oximetry sensor to
the patient correctly
157).
Insert battery
Reset date
If this recurs: Have the device
repaired.
Have the device repaired.
Insert battery correctly
83).
Insert new SD card
92), switch off device
119), and switch device back on
after at least 30 s
To continue to record the session:
Insert SD card
copy data to SD card, and resume
session with SD card.
Resume session and do not record
any more data.
Connect SpO
pulse oximetry sensor connecting
cable to the device correctly.
Change printer battery
p.
97).
Charge printer battery
p.
94).
MEDUCORE Standard
(see 6.9.1, p.
(see 4.3, p.
83).
(see 10.3.7, p.
264).
(see 4.3, p.
(see 4.6.1, p.
(see 6.2, p.
(see 6.1, p.
118).
(see 4.6.1, p.
92),
connector of the
2
(see 4.7.2,
(see 4.7.1,
2
EN
277