Risk of injury due to incorrect handling of electrode gel!
Incorrect handling of electrode gel may lead to electric shock, to
ineffective shock delivery, and to burns, injuring the patient, user,
and bystanders.
 Always use electrode gel with paddles.
 Do not allow any electrode gel to get between the surface of
the electrode and the handle in order to prevent electric shock.
 Do not use too much electrode gel to prevent a gel bridge and
thus burns on the chest.
 Do not use too little electrode gel to keep the resistance to the
patient low for effective shock delivery and to prevent burns on
the chest.
Risk of injury from the incorrect electrode gel!
The incorrect electrode gel may trigger intolerance reactions and
lead to ineffective shock delivery. This may injure the patient.
 Only use electrode gel recommended by
WEINMANN Emergency.
7. Wet the electrode surfaces of the paddles completely with
electrode gel.
Result
Shock delivery via paddles is prepared.