Weinmann MEDUCORE Standard2 Instructions For Use Manual page 139

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

It is not possible to switch to parameter view in manual
mode. If you activate manual mode from parameter view,
the device automatically switches to curve view, since for
manual shock delivery the ECG analysis is required in the
display.
Audio alarm output is deactivated in manual mode.
3. To activate audio alarm output: Briefly press the alarm button
.
4. Evaluate the ECG lead.
5. If shock is required: Select shock energy using the Energy
function button.
When doing so, please note: When the defibrillation electrodes
for children (Pediatric) are connected, the shock energy is
automatically restricted to 100 J. It is not possible to set a
higher shock energy in manual mode.
6. Press the Charge function button.
The charge progress bar appears. A rising charging tone
sounds until the device is ready to deliver the shock. When the
device is charged, a sequence of tones sounds which signals
shock standby and the shock button
7. To cancel shock charging: Cancel shock charging by pressing
the Cancel function button or by switching to another mode.
8. Check the ECG leads to see whether defibrillation is still
indicated.
Risk of injury from electric shock!
The electric shock administered to the patient may injure the user
or bystanders.
 Stand clear of the patient.
 Keep patient away from liquids (e.g. blood, gel or saline
solution).
 Do not touch parts in contact with the patient (e.g. bed frames
or stretchers).
 Keep your distance from liquids in contact with the patient.
 Clearly warn bystanders to stand clear of the patient or parts in
contact with the patient and to keep their distance from liquids
in contact with the patient.
flashes.
MEDUCORE Standard
6 Operation
2
EN
139