Weinmann MEDUMAT Easy CPR Description And Instructions For Use

Weinmann MEDUMAT Easy CPR Description And Instructions For Use

Ventilator
Hide thumbs Also See for MEDUMAT Easy CPR:
Table of Contents

Advertisement

Quick Links

MEDUMAT Easy CPR
ERC 2010
Ventilator
Description and instructions for use

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MEDUMAT Easy CPR and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Weinmann MEDUMAT Easy CPR

  • Page 1 MEDUMAT Easy CPR ERC 2010 Ventilator Description and instructions for use...
  • Page 2: Table Of Contents

    Warranty ....73 MEDUMAT Easy CPR ..42 Declaration of conformity ..73...
  • Page 3: Overview

    1. Overview 1.1 Device MEDUMAT Easy CPR control panel 2 Ventilation pressure gauge 3 Alarm panel MEDUMAT Easy CPR mbar Stenosis 4 Alarm mute button Disconnection < 2,7 bar O 5 Colour code 6 CPR key and LED 7 Regulator knob, ventilation...
  • Page 4 13 Speaker 14 MEDUtrigger connection 15 Pressure gauge hose connection 16 Ventilation hose connection 17 Relief outlet valve MEDUMAT Easy CPR device combinations 18 Ventilation mask C P R Disposable hose system 19 Ventilation hose 20 Filter 21 Pressure gauge hose...
  • Page 5: Particular Markings On The Device

    1.2 Particular markings on the device Patient valve symbol on the patient valve is a reminder that the lip membrane must be changed immediate- ly if it becomes crinkled, sticky or misshapen. The pa- tient must not be used again until the membrane has been changed, as this could cause malfunction (see “7.4 Checking patient hose system”...
  • Page 6 Inlet 2.7 - 6 bar O RU, PL, CS, ZH, JA, AR, TR Languages available on this device ID, TH, FA, DE, EN, KO, HI MEDUMAT Easy CPR ID plate Serial number of device Date of manufacture 3.6 V lithium battery...
  • Page 7 Follow instructions for use Tube system connection Maximum pressure ≤100 mbar Indicates where to connect the MEDUtrigger. Safety check and servicing label Safety check [STK] label: (Germany only): indicates when the next safety check as per §6 of the German law governing medical devices and their owners/operators is required.
  • Page 8 MEDUtrigger MEDUtrigger ID plate Protection class BF Do not dispose of the device in domestic waste CE symbol (confirms that the product conforms to the applicable European directives) IP54 Protection against ingress of dust and water Protection class II, protective insulation Date of manufacture Keep connector upright and do not rotate it when taking it out EN/HI...
  • Page 9: Description Of Device

    2. Description of device 2.1 Intended use MEDUMAT Easy CPR is an automatic oxygen respira- tion device (short-term ventilator) with additional in- halation facility. You can use MEDUMAT Easy CPR: • to revive patients at the site of the emergency;...
  • Page 10: Owner/Operator And User Qualification

    2.3 Ventilation function MEDUMAT Easy CPR operates within a pressure range of 2.7 to 6 bar and at a flow rate of not less than 70 l/min O . It has a built-in power pack.
  • Page 11: Demandflow Function

    2.4 Demandflow function The Demandflow setting switches the Demandflow setting MEDUMAT Easy CPR breathing-controlled O inha- lation. Such inhalation must be carried out with the respiration mask. A small inspiration (trigger-) pulse causes oxygen to continue flowing until slight over- pressure interrupts the flow.
  • Page 12: Audio Response

    2.6 Audio response The device has an audio response facility that can be switched on for user guidance, especially for users who have little practice. If audio response is not required, a key combination can be used to switch it off (see “5.11 Audio re- sponse for user guidance”...
  • Page 13: Safety Instructions

    (see “7.4 Checking patient hose system” on page 50). Note • Do not use the MEDUMAT Easy CPR in toxic en- vironments or where there is a risk of explosion. • MEDUMAT Easy CPR is not suitable for hyper- baric use (pressure chamber).
  • Page 14 To prevent infection or bacterial contamination, please observe section “6. Hygiene preparation” on page 42. • MEDUMAT Easy CPR should be used only by medically qualified personnel who have had training in ventilation techniques. Severe physi- cal damage may be caused by incorrect use.
  • Page 15 • Smoking and open flames are strictly prohibited in the vicinity of all fittings containing or trans- porting oxygen. • During assembly and when changing oxygen bottles only hand-tighten screwed connections to the bottle and the pressure reducer. Never use tools for this purpose.
  • Page 16 Accessories • Keep silicone/rubber parts out of UV light and avoid prolonged exposure to direct sunlight, otherwise these materials may become brittle and crumble. EN/HI Safety instructions...
  • Page 17: Installation

    4. Installation Permanent mounting is usually necessary only when MEDUMAT Easy CPR is installed as a fixture in rescue vehicles, helicopters or aircraft. If MEDUMAT Easy CPR is supplied complete on a carrying platform or in an emergency rucksack, it is ready for use and requires no further installation.
  • Page 18: Ventilation Hose And Medutrigger

    G 3/8. 4. Screw the other end of the pressure hose on to pressure gas connection on the MEDUMAT Easy CPR if this has not yet been done. CP R 4.2 Ventilation hose and MEDUtrigger 1.
  • Page 19 connecting or disconnecting it. The connection socket may be damaged if the MEDUtrigger connector is rotated. 2. Slide the pressure gauge tube onto connection. 3. Slide the ventilation hose onto connection. C P R Make sure that this does not cause any kinks in the pressure gauge tube already connected.
  • Page 20 if the patient is intubated, attach the patient valve to the tube. Filter If a filter is to be used, this should be installed be- tween the patient side connector on the patient valve and the mask or tube. In this case, the MEDUtrigger is attached to the filter.
  • Page 21: Wall Mounting Set

    4.3 Wall mounting set A wall mounting set (see “10.2 Accessories” on page 66) is available for permanent fixing, e.g. on a vertical surface inside a vehicle. Please refer to the sheet enclosed with the wall mounting set for details of dimensions and installa- tion procedure.
  • Page 22: Using The Ventilator

    3. Select the desired ventilation settings (see “5.2 Selecting the ventilation settings” on page 23). 4. Switch on the MEDUMAT Easy CPR with ON/ OFF switch. The ventilator will then run a self- test lasting approx. 2 seconds.
  • Page 23: Selecting The Ventilation Settings

    Otherwise an alarm is sounded. The MEDUMAT Easy CPR will then start to func- tion with the selected ventilation settings. 5.2 Selecting the ventilation settings We recommend selecting the ventilation settings be- fore switching on, to prevent unnecessary waste of oxygen.
  • Page 24: Performing Ventilation

    45 mbar”) If the maximum ventilation pressure is reached, e.g. in cases where compliance is inadequate, MEDUMAT Easy CPR sets off a stenosis alarm (see “Stenosis alarm” on page 31). Note: The mask/tube switch can only be operated when the ventilator is switched on.
  • Page 25: Monitoring Ventilation

    Intubation The patient will normally be intubated before the tube is connected to the patient valve. Risk of injury from the tube moving! The MEDUtrigger on the patient valve may move the tube and injure the patient. Detach the MEDUtrigger from the patient valve before putting the patient valve on the tube.
  • Page 26: Ventilation With Peep Valve

    If lung compliance diminishes during ventilation, the ventilator will react by increasing the ventilation pressure at constant volume. Example of ventilation sequence before and after compliance diminishes 5.5 Ventilation with PEEP valve A PEEP valve can be fitted to the expiration tube on the patient valve with an adapter.
  • Page 27: Demandflow

    Demandflow mode! Demandflow must be switched on for O inhalation. To switch the MEDUMAT Easy CPR to Demandflow mode, turn the ventilation settings regulator knob until it engages in the fixed point marked by the white triangle. The green LED indicates that it is Freq.(min...
  • Page 28: Cpr Mode

    5.8 CPR mode Automatic ventilation with the selected frequency is stopped when CPR mode is activated. With the help of MEDUtrigger, you can trigger single respiratory strokes with the selected tidal volume. This enables you to choose the respiratory frequency adminis- tered yourself.
  • Page 29 (see “5.11 Audio response for user guidance” on page 34 and “5.12 Switching the metronome on and off” on page 38). This makes particular sense if the MEDUMAT Easy CPR is being used with other devices that have audio responses and/or a metronome (e.g. AED).
  • Page 30: Terminating Ventilation Or Demandflow

    This prevents entry of moist atmospheric air that can cause corrosion. 3. Switch off MEDUMAT Easy CPR. To prevent the ventilator being switched off unintentionally, ON/ OFF switch must be kept pressed down for at least 2 seconds until the LEDs in the alarm panel light CP R up.
  • Page 31 “ डवाइस म खराबी है ! वै क ि पक व टले श न दान कर” (“Device malfunction! Administer alternative ventilation!”). In this case you must not use the MEDUMAT Easy CPR. The failure alarm can be confirmed by pressing the ON/ OFF switch. The patient valve is designed to enable spontaneous CP R breathing in the event of equipment failure.
  • Page 32 If the maximum ventilation pressure is exceeded for a second time during the same inspiration phase, the unit finally switches to expiration and vents the pa- tient tube system completely. The next inspiration begins with the following ventilation stroke accord- ing to the frequency selected.
  • Page 33 < 2.7 bar O alarm Oxygen pressure at the pressure connection to the MEDUMAT Easy CPR has dropped to less than 2.7 bar. The reason is usually an almost empty oxygen cylinder. In this case, MEDUMAT Easy CPR can no longer function correctly because the operating parameters are no longer within the permissible limits.
  • Page 34: Audio Response For User Guidance

    The ventilator must be switched off before the battery can be changed (see “Changing the main battery” on page 63). Cancelling acoustic alarm If there is an alarm, you can temporarily cancel the acoustic alarm by pressing the alarm mute button: Stenosis: 30 seconds Disconnection:...
  • Page 35 language setting. The following languages are assigned to the individual LEDs: Language, Level 2 Device number mbar Language, Level 1 (Stenosis and disconnection alarm LEDs come on) Icelandic Finnish Norwegian Swedish Danish Still free Portuguese WM 28140 WM 28150 Spanish (French variant) Dutch Italian...
  • Page 36 3. Press mask/tube switch as many times as neces- sary until the LED for the desired language lights up and a corresponding message is heard (exam- ple: LED 10 mbar, language: English, message: “Selected language: English.”). After five sec- onds the new selection is stored. CP R By pressing the ON/OFF switch briefly you can store the language selection without waiting for...
  • Page 37 (page 41). ventilation!”) “ े श र क नल तथा गै स स लाई MEDUMAT Easy CPR has measured low pressure on the inlet side. जाँ च ल” (“Check pressure hose Check whether the O cylinder still contains sufficient oxygen and system and gas supply!”)
  • Page 38: Switching The Metronome On

    MEDUtrigger at the right time. Ventilation triggered manually!”) “सीपीआर मोड नि य है ! ” MEDUMAT Easy CPR ventilates with the selected frequency. (“CPR mode deactivated!”) “दो व टले श न दान कर!” Trigger 2 respiratory strokes with the MEDUtrigger. (“Provide 2 ventilations now!”) “अब...
  • Page 39: Calculation Of Oxygen Content/Remaining Operating Time

    – 2 x audible confirmation: metronome activation is confirmed 5.13 Calculation of oxygen content/remaining operating time Oxygen content of cylinder Oxygen volume = volume of cylinder x cylinder pressure. Cylinder volume x cylinder pressure = oxygen content Example 1 10 l x 200 bar = 2000 l Example 2...
  • Page 40 Real Demandflow time Example: Max. Flow Inhalation flow 45 l/min = 1.5 l = 1.5 l Patient data: Inspiration: Expiration (I : E) = 1 : 2 I = 2 sec E = 4 sec I = 2 sec Respiratory frequency = 10 min = 0.033 min = 0.066 min = 0.033 min...
  • Page 41: Alternative Ventilation Procedures

    5.14 Alternative ventilation procedures If MEDUMAT Easy CPR ceases to function during a ventilation procedure, the following alternatives can be applied: Ventilation bags 1. Remove the patient valve from the tube or the mask. 2. Replace it with the ventilation bag, e.g. a WEINMANN Emergency COMBIBAG WM 11000, and perform manual ventilation.
  • Page 42: Hygiene Preparation

    6.1 MEDUMAT Easy CPR You can keep MEDUMAT Easy CPR clean by simply wiping with disinfectant as described in section 6.7. Never immerse MEDUMAT Easy CPR in disinfectants or other liquids.
  • Page 43: Ventilation Hose

    2. Dismantle the patient valve as shown in the ad- jacent diagram. It is neither necessary nor per- missible to remove the membrane in the spontaneous breathing tube for cleaning and disinfection. membrane Disc Disc membrane membrane Expiration Spontaneous tube breathing tube 3.
  • Page 44: Medutrigger

    1. Take the ventilation hose and the pressure gauge hose off bothconnection ports. C P R Caution! at the end Always grasp the hoses , as shown in the drawing, otherwise the hoses may be damaged or torn off. 2. Pull the MEDUtrigger plug down and out with- out turning or levering it (see diagram).
  • Page 45: Method

    6.7 Method Hygienic preparation of the MEDUMAT Easy CPR and the accessories used should be performed as de- scribed in the following table. Follow the instructions for use for the disinfectant ®...
  • Page 46 (1) If disinfecting by wiping is required: ensure that no liquids pe- netrate the connections. Cleaning agents containing alcohol or moisturizers form a flammable mixture in combination with compressed oxygen and may lead to explosions. (2) To disinfect the pressure gauge tube of the ventilation hose, proceed as follows: 1.
  • Page 47: Functional Check

    7. Functional check MEDUMAT Easy CPR must not be used if the func- tional checks reveal defects or deviations from the specified parameters. First try to correct the error with the help of the in- formation provided in section “8. Troubleshooting”...
  • Page 48: Intervals

    Note Check the test bag before each functional check. The balloon of the test bag must be un- damaged and firmly connected to the connec- tor. Have the test bag serviced at the same time as the device. 7.2 Intervals Before each use: •...
  • Page 49: Checking For Leaks In The System

    Bubbles will form at the site of the leak. 3. Depressurize the system: To do this, first close the oxygen cylinder. Switch on MEDUMAT Easy CPR briefly until the pres- sure gauge on the O cylinder reads “0”. Then switch off MEDUMAT Easy CPR again.
  • Page 50: Checking Patient Hose System

    7.4 Checking patient hose system Checking reusable hose system 1. Dismantle the patient valve. 2. Carry out a visual check of all components for cracks or other physical damage. The lip membrane must be replaced if it is crin- kled, sticky or misshapen. It must no longer be used for ventilation as it could cause serious functional problems.
  • Page 51: Checking The Tidal Volume

    5. Turn up the frequency to 10 min (right limit before the index point). 6. Count the number of inspiration phases over a period of exactly one minute. The figure should lie between 8 and 12. 7. Switch off MEDUMAT Easy CPR again. EN/HI Functional check...
  • Page 52 Checking the tidal volume 1. MEDUMAT Easy CPR must be switched off and the oxygen cylinder must be open. 2. Attach the test bag to the patient valve with the adapter from test set WM 15357. 3. Select the following settings: –...
  • Page 53: Checking Maximum Ventilation Pressure

    – : 65 ml / frequency 25 min – (45 mbar) 8. Switch on MEDUMAT Easy CPR and close the patient connection at the patient valve. A stenosis alarm should be triggered. 9. Switch off MEDUMAT Easy CPR again. 7.6 Checking maximum ventilation pressure 1.
  • Page 54: Test Demand Flow

    8. Detach the test bag from the patient valve. 7.7 Test demand flow 1. Make sure that MEDUMAT Easy CPR is switched off and the oxygen cylinder is open. 2. Attach the test bag to the patient valve with the adapter from test set WM 15357.
  • Page 55: Checking The Alarm Systems

    7. Switch off MEDUMAT Easy CPR again. Risk of injury if test bag removed incorrectly! If the test bag is removed incorrectly, the con- nector of the test bag may remain on the pa- tient’s hose system. The increased airway resis- tance on inspiration which results may injure the patient.
  • Page 56 If audio response is enabled, the ventilator announces “व टले श न स टम और से टं स क जां च कर!” (“Check ventilation system and settings!”). 7. Switch off MEDUMAT Easy CPR again. Drop in O pressure (<2.7 bar O slowly 1.
  • Page 57: Checking The Medutrigger

    MEDUMAT Easy CPR is switched on. The power supply is in order if no alarm is triggered when the valve on the oxygen cylinder is opened and MEDUMAT Easy CPR is switched on and starts to function. 7.9 Checking the MEDUtrigger 1.
  • Page 58 Risk of injury if test bag removed incorrectly! If the test bag is removed incorrectly, the con- nector of the test bag may remain on the pa- tient’s hose system. The increased airway resis- tance on inspiration which results may injure the patient.
  • Page 59: Troubleshooting

    8. Troubleshooting Fault Cause of fault Remedy MEDUMAT Easy CPR defective. Have repaired. Replace battery in battery compartment (9.3, page 62). MEDUMAT Easy CPR will If ventilator still refuses to switch on, not switch on. Battery failing. have internal auxiliary battery replaced by manufacturer or authorized specialists.
  • Page 60 Activate audio response (5.11, No audio response. Audio response deactivated. page 34). Pressure gauge hose on MEDUMAT Easy CPR is MEDUMAT Easy CPR or on functioning, but without Check pressure gauge hose. patient valve slipped off. any displays. Kink in pressure gauge hose.
  • Page 61: Servicing

    9. Servicing 9.1 Intervals Have the cleaned and disinfected device serviced at regular intervals. Servicing, safety checks ([sicherhe- itstechnische Kontrollen or STKs] in accordance with §6 of the German law governing the owners/opera- tors of medical devices - only applies to Germany) and maintenance measures such as servicing and re- pairs may only be performed by the manufacturer or by specialists expressly so authorized by the manu-...
  • Page 62: Sending In Device

    WEINMANN Emergency at your expense. 9.3 Batteries MEDUMAT Easy CPR is equipped with two batteries: • Main battery (lithium battery 3.6 V) for main power supply. This may be changed by the operator.
  • Page 63 Changing the main battery 1. Make sure the ventilator is switched off. 2. Open the battery compartment on the side of the MEDUMAT Easy CPR (e.g. with a coin). 3. Remove the old 3.6 V lithium battery. C P R 4.
  • Page 64: Change Valve Membrane In Patient Valve

    2. Put in a new valve membrane. 9.5 Storage If you do not intend to use MEDUMAT Easy CPR for a long period, we recommend the following storage precautions: 1. Clean and disinfect (see “6. Hygiene prepara- tion”...
  • Page 65: Disposal

    Disposing of batteries/rechargeable batteries Used batteries/rechargeable batteries may not be disposed of in domestic waste. Contact WEINMANN Emergency or your local authority waste disposal de- partment. Disposing of the patient hose system After use, the patient hose system should be disposed of properly with plastics.
  • Page 66: Scope Of Supply

    10. Scope of supply 10.1 Standard scope of supply MEDUMAT Easy CPR, complete 28140 consisting of: – MEDUMAT Easy CPR, single unit 28237 – Operating instructions 67281 – Summary of operating instructions 67291 – Set of mounting attachments 15007 –...
  • Page 67: Replacement Parts

    15. Respiratory system filter 22162 10.3 Replacement parts You can order replacement parts separately if required. You can obtain a current list of replacement parts either on the Internet at www.weinmann-emergency.com or through your specialist dealer. EN/HI Scope of supply...
  • Page 68: Technical Data

    11. Technical data 11.1 Device MEDUMAT Easy CPR MEDUMAT Easy CPR 1:2 (V ≤ 150 ml), Dimensions 100x145x90 Insp-exp. ratio (I:E) 1:3 (V > 150 ml) WxHxD in mm incl. connections Respiratory time ratio Weight incl. approx. 0.6 kg (I:E) in CPR mode...
  • Page 69 MEDUMAT Easy CPR MEDUMAT Easy CPR Pressure gauge ±5% of upper range Power supply maintenance-free lithium accuracy value battery 3.6 V; 5.2 Ah, Life expectancy > 2 years Alarm sound 60 dB (A) Max. storage period 10 years after delivery...
  • Page 70: Patient's Hose System

    22 mm external tapered connector – mask/endotracheal tube EN 5356-1 Patient valve 30 mm external tapered connector – expiration tube EN 5356-1 Ventilation tube connection WEINMANN Emergency-specific Standards applied EN 794-3 Patient valve resistance (as per EN 794-3): Inspiration <6 mbar at 60 l/min <6mbar...
  • Page 71: Pneumatic / Electronic Systems

    11.3 Pneumatic / electronic systems Description of function 3/2-way valve Relief valve Pneumatic 100mbar Patient valve Ventilation Pressure regula- hose RF/MV 2.7-6bar 2.5bar Exp. Flow Booster valve Pressure Coupling element with metering facili- measuring Insp. hose Respiratory frequen- Demandflow Electronic setting system Battery...
  • Page 72: Relationship Between Ventilation

    Demandflow An inspiration impulse (trigger) at V2 opens valves V3 and V4. Electronic system The microprocessor-controlled electronic system sets the ventilation parameters and monitors ventilation, and also O supply and power supply. If necessary, a visual and acoustic alarm is given. The ventilator has an audio response facility that can be switched on for user guidance.
  • Page 73: Warranty

    12. Warranty WEINMANN Emergency gives the customer a limited manufacturer warranty on new original WEINMANN Emergency products and any replacement part fitted by WEINMANN Emergency in accordance with the warranty conditions applicable to the product in ques- tion and in accordance with the warranty periods from date of purchase as listed below.
  • Page 76 Manufacturer Center for Production, Logistics, Service WEINMANN Emergency WEINMANN Emergency Medical Technology GmbH + Co. KG Medical Technology GmbH + Co. KG Frohbösestraße 12 Siebenstücken 14 22525 Hamburg 24558 Henstedt-Ulzburg GERMANY GERMANY T: +49 40 88 18 96-120 E: customerservice@weinmann-emt.de...

Table of Contents