13.4.2 Possible values
Parameter
Heart rate
Heart rate
Pulse rate
Pulse rate
SpO
SpO
Adult alarm limit
settings
NIBP systolic
NIBP systolic
NIBP diastolic
NIBP diastolic
Possible
values
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
35/min - 250/min,
in increments of 5
30/min - 245/min,
in increments of 5
saturation
66 % - 100 %
2
saturation
65 % - 99 %
2
45 mmHg -
260 mmHg,
in increments of 5
40 mmHg -
255 mmHg,
in increments of 5
25 mmHg -
200 mmHg,
in increments of 5
20 mmHg -
195 mmHg,
in increments of 5
13 Operator menu
Description
Here you can set the upper
( ) and lower ( ) limit
values for the patient group
as of which the device is to
emit an alarm.
According to the values set,
the setting ranges of the
upper and lower limit values
depend on each other as
follows:
The setting range for the
upper alarm limit ends at
least one setting value
above the set value for
the lower alarm limit.
The setting range for the
lower alarm limit ends at
least one setting value
below the set value for
the upper alarm limit.
MEDUCORE Standard
Factory
setting
120/min
50/min
120/min
50/min
100 %
85 %
220 mmHg
75 mmHg
110 mmHg
35 mmHg
2
EN
317