Weinmann MEDUCORE Standard2 Instructions For Use Manual page 243

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Manual
cleaning
(only
Part
necessary in
the event of
visible dirt)
Pulse oximetry monitoring
Wipe down with
neodisher
MediClean forte
MCS2-Adapt
(Dr. Weigert).
pulse oximetry
Dose:
sensor
10 ml/l
connecting
cable
Wipe down all
surfaces at least
2x until they
look clean.
MCS2-SoftTip
pulse oximetry
Wipe down with
sensor
neodisher
MCS2-Wrap
MediClean forte
pulse oximetry
(Dr. Weigert).
sensor
Dose:
10 ml/l
MCS2-Earclip
pulse oximetry
Wipe down all
sensor
surfaces at least
Ear clip for
2x until they
MCS2-Earclip
look clean.
pulse oximetry
sensor
MCS2-Wrap
pulse oximetry
sensors
(disposable)
Strap for
Not permitted, as disposables
attaching
MCS2-Wrap
pulse oximetry
sensor
(disposable)
Disinfection
by wiping
®
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
®
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
Disinfection
Mechanical
by
reprocessing
immersion
Not permitted
Not permitted
Not permitted
Not permitted
MEDUCORE Standard
8 Hygienic reprocessing
Sterilization
Not permitted
Not permitted
2
EN
243