Alarm
NIBP signal disrupted
(Alarm deactivated after 10 s)
NIBP systolic
(Alarm deactivated after 10 s)
NIBP systolic
(Alarm deactivated after 10 s)
NIBP tube blocked
(Alarm deactivated after 10 s)
NIBP tube leaking
(Alarm deactivated after 10 s)
No printer connected
(Alarm deactivated after 10 s)
Plug in ECG connector
Plug in Pad connector
(Monitor mode)
Cause
Unable to detect pulse wave signal
for NIBP measurement reliably
Measured systolic blood pressure
is above the set alarm limit
Measured systolic blood pressure
is below the set alarm limit
NIBP connecting tube kinked or
blocked
NIBP connecting tube or NIBP cuff
are leaking or incorrectly
connected. Sufficient pressure
cannot be built up.
Connection between device and
printer interrupted
ECG connector of the ECG cable
not connected/not connected
correctly to the device
Defibrillation electrodes not
connected/not connected correctly
to the master cable
Paddles not connected/not
connected correctly to the master
cable
Master cable not connected/not
connected correctly to the device
11 Alarms and faults
Remedy
Make sure that the patient lies
still during the measurement.
Rule out the risk of patient
vibration.
Use a suitable size of NIBP
cuff.
Ensure that the NIBP cuff is
properly attached.
Ensure that there is no
clothing between the NIBP
cuff and the patient.
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
255).
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
255).
Check NIBP connecting tube and
replace if necessary.
Check NIBP connecting tube and
NIBP cuff and replace if necessary.
Wait 10 seconds.
Switch printer on again.
Position the printer within range of
the device.
Connect ECG connector of the
ECG cable to the device correctly
(see 6.10.1, p.
Connect defibrillation electrodes
to the master cable correctly
6.5.1, p.
122).
Connect paddles to the master
cable correctly
Connect the master cable to the
device correctly
MEDUCORE Standard
(see 10.3.1, p.
(see 10.3.1, p.
164).
(see
(see 6.5.2, p.
126).
(see 6.5.1, p.
122).
2
EN
275