9. Maintenance

9.1 Intervals

We recommend only having servicing and maintenance performed by the
manufacturer (WEINMANN) or authorized dealers.
Both filters must be regularly checked for soiling
The coarse dust filter should be cleaned once a week and changed
no less than every 6 months.
The fine filter should be changed after no more than 1000 operat-
ing hours (filter change symbol
For reasons of hygiene, we recommend replacing the following parts at the intervals
stated:
pressure measuring hose every 6 months and earlier if soiled
complete mask system every 6 to 12 months depending on soiling
exhalation system in accordance with the relevant operating
instructions
In addition, servicing should be carried out at the following intervals as a preventive
measure:
after every 5000 operating hours (maintenance symbol
appears in the display)
after no more than 2 years (see servicing sticker on rear of unit)
Servicing includes:
filter change
cleaning unit
replacing any defective parts
a complete check of unit functions and pressure indicators
appears in the display).
Maintenance
39
EN