Weinmann MEDUCORE Standard2 Instructions For Use Manual page 242

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

8 Hygienic reprocessing
Manual
cleaning
(only
Part
necessary in
the event of
visible dirt)
Defibrillation/cardioversion
MCS2-Connect
Wipe down with
master cable
neodisher
MediClean forte
MCS2-Hardpads
(Dr. Weigert).
paddles
Dose:
10 ml/l
Function test
Wipe down all
resistor
surfaces at least
2x until they
look clean.
MCS2-Softpads
defibrillation
electrodes for
children/adults
Not permitted, as disposables
(disposable)
Electrode gel
(disposable)
242
EN
MEDUCORE Standard
Disinfection
by wiping
®
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
2
Disinfection
Mechanical
by
reprocessing
immersion
Not permitted
Not permitted
Sterilization
Not permitted

Advertisement

Table of Contents
loading

Table of Contents