6.20 Archiving session data

From 60 seconds after the start of a session, the device
permanently saves session data to its internal memory and on the
SD card if the SD card is in the SD card slot at the beginning of the
session. After a session has started, session data are saved only
temporarily in the internal memory and not saved on the SD card
and are deleted if you switch off the device in < 60 s.
If you switch off the device and then switch it on again after < 30 s,
the device saves session data in the session data set already started.
If you switch off the device and then switch it on again after > 30 s,
the device creates a new session data set.
6.20.1 Exporting session data from the internal
memory to SD card
Requirement
There is an SD card in the SD card slot.
1. Activate the operator menu
2. Select System settings | SD card | Export internal memory
3. Confirm with the OK function button.
4. Remove the SD card
Result
Session data and service data have been exported from the internal
memory to the SD card.
menu", page
307).
to SD card.
The export process starts.
page
93).
(see "13.1 Activating the operator
(see "4.6.2 Removing the SD card",
MEDUCORE Standard