4 Preparation
Result
4.4.4 Changing the battery
Requirement
Result
88
EN
MEDUCORE Standard
Battery not fully charged (< 95 % charge)
Battery temperature between 0 °C and 45 °C
The battery status LEDs flash green during the charging
process.
2. Once the status LEDs go permanently green: Release the
battery latch and take the battery out of the charging station.
The battery is fully charged.
The replacement battery is fully charged.
1. If the device is not connected to line power: Switch off the
device
(see "6.2 Switching the device off", page
2. Take battery out of the battery compartment.
Risk of injury from battery being incorrectly inserted!
An incorrectly inserted battery may lead to device failure and thus
injure the patient.
 Always push the battery into the battery compartment until
you hear it engage and it is flush with the housing.
3. Slide the replacement battery into the battery compartment
until you hear it engage.
4. If necessary: Switch on the device
device", page
118).
The symbol
The battery is changed.
2
(see "6.1 Switching on the
appears in the display.
119).