2 Safety
Caution
20
EN
MEDUCORE Standard
Impaired readiness for use as a result of a defective power
cord or defective power supply unit and charger!
A defective power cord or a defective power supply unit and
charger prevent the battery charging in the charging station and
thus the readiness of devices operated with the battery.
 Check the power cord and the power supply unit and charger
regularly.
Risk of injury from electric shock!
The contacts in the battery compartment of the charging station
are live. Touching the contacts may lead to electric shock and
injure the user.
 Do not touch the contacts in the battery compartment of the
charging station.
Risk of injury from touching the contacts in the battery
compartment and the patient at the same time!
The contacts in the battery compartment are live. Touching the
contacts and the patient at the same time may injure the user or
the patient.
 Only operate the device with the battery inserted.
Risk of injury due to trailing power cord!
Trailing power cords are a trip hazard, which may hinder operation
and cause injuries.
 During operation with line power, position the power cord so
that it does not present a hindrance.
 During 12 V operation, position the power cord so that it does
not present a hindrance.
Risk of injury as a result of falsified measurement results
when the line filter is not set correctly!
An incorrectly set line filter may falsify the measurement results of
the ECG and lead to misinterpretations. This may injure the
patient.
 Adapt the line filter to suit the regional power supply network.
 When assessing the ECG, take account of an activated line
filter.