11-1 Alarm line with VF/VT alarm (example)
The device displays alarms as follows:
As text in the alarm line in the display
As audio alarm tones (via the loudspeaker)
With the alarm light (in the top right-hand corner on the front
of the device)
Alarms are outputted according to their priority:
Type of alarm
High priority
signal
Alarm line
Flashing
2 Hz
frequency
Size
90 mm x 5 mm
Color
Red
Signal ratio
1:1
11 Alarms and faults
Medium
priority
0.5 Hz
Yellow
1:1
MEDUCORE Standard
Low priority
Not flashing
(constantly on)
Turquoise
100 % (on)