Weinmann MEDUCORE Standard2 Instructions For Use Manual page 191

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Risk of injury due to incorrectly selected patient group!
The device only delivers correct measured values if the appropriate
patient group is selected. An incorrect patient group may lead to
incorrect measurements and injure the patient.
 Adapt the patient group to the patient.
 If the patient group is incorrect: Change the patient group in
the application menu.
2. If the patient group is incorrect: Select another patient group
(see "6.4 Selecting patient group", page
The NIBP module is configured accordingly in the device with
the selected patient group.
3. Press the Interval function button.
4. Set the interval duration using the navigation knob.
5. Press the Start function button.
Alternatively:
Press the NIBP button
An NIBP measurement starts. After the end of the
measurement, the device displays the systolic and diastolic
arterial blood pressure, the timer counts down, and when it has
finished the next measurement starts automatically.
6. Evaluate the NIBP measurement result.
7. If necessary: Cancel the NIBP measurement with the Stop
function button.
The device releases the pressure from the NIBP cuff.
8. If necessary: Set alarm limits in the user menu
Alarm settings", page
Alternatively:
Set automatic alarm limits in the application menu
Application menu", page
120).
for > 2 s.
255).
248).
MEDUCORE Standard
6 Operation
(see "10.3.1
(see "9
2
EN
191