6 Operation
Result
6.5.2 Preparing for shock delivery using paddles
126
EN
MEDUCORE Standard
12. Attach defibrillation electrodes and press in place firmly.
Risk of injury from defibrillation electrodes not adhering
correctly!
If defibrillation electrodes have been attached incorrectly and are
attached again, they no longer adhere correctly and may lead to
inadequate shock delivery. This may injure the patient.
 Always keep spare defibrillation electrodes to hand.
 If defibrillation electrodes are attached incorrectly: Always use
new defibrillation electrodes.
13. If defibrillation electrodes are attached incorrectly: Use new
defibrillation electrodes instead of the old defibrillation
electrodes.
14. Stroke out any air trapped under the defibrillation electrodes.
Shock delivery via defibrillation electrodes is prepared.
Risk of injury from using paddles with the incorrect software
version!
On devices with a software version < 2.1, the device does not
detect paddles so they cannot be used correctly. This may injure
the patient and the user.
 Only use paddles on devices running software version 2.1 or
above.
1. Connect the master cable connector to the Pad connection for
master cable on the device.