13 Operator menu
13.3.2 Possible values
Parameter
Adult AED settings/Child AED settings
First shock
Energy
progression
Further shocks
Duration
CPR phase
Pause for
ventilation
312
EN
MEDUCORE Standard
Possible
Description
values
Here you can set the shock
energy of the first shock in
AED mode.
1 J to 200 J
Here you can set the shock
energy of further shocks in
AED mode.
Here you can set the time
interval between cardiac
rhythm analyses.
The device only outputs
120 s to 300 s
certain AED instructions and
voice prompts after CPR is
complete, as complete and
correct carrying out of CPR
takes priority.
Here you can set the time
interval for ventilation
between chest compressions.
When choosing a setting,
consider whether voice
prompts are deactivated:
3 s to 8 s
2
When voice prompts are
activated, the pause
begins after the last
metronome sound and
ends when the voice
prompts begin.
When voice prompts are
deactivated, the pause
begins after the last
metronome sound and
ends when the first
metronome sound
begins.
Factory
setting
AED mode
adult: 150 J
AED mode
child: 75 J
AED mode
adult: 200 J
AED mode
child: 75 J
120 s
5 s