Alarm
ECG module defective
E-mail delivery error
(Alarm deactivated after 60 s)
Excessive NIBP cuff pressure
(Alarm deactivated after 10 s)
Heart rate
Heart rate > 250/min
Heart rate
Heart rate < 30/min
Insert printer paper
Memory error
NIBP diastolic
(Alarm deactivated after 10 s)
Cause
Internal ECG module defective
Server not accessible
Connection between device and
network interrupted
Pressure in the NIBP cuff increased
to such a degree as a result of
external influences that the safety
valve opened
Measured heart rate is above the
set upper alarm limit
Measured heart rate is above 250/
min (device shows --- as heart
rate)
Measured heart rate is below the
set lower alarm limit
Measured heart rate is below 30/
min (device shows --- as heart
rate)
Printer paper completely used up
No printer paper inserted
Internal memory module defective Have the device repaired.
Measured diastolic blood pressure
is above the set upper alarm limit
11 Alarms and faults
Remedy
Have the device repaired.
Check connection to network
12.3, p.
292).
Check preset e-mail recipient
12.5, p.
299).
Check connection to network
12.3, p.
292). If necessary: Select a
different network.
Check date/time
264).
Check which influences resulted in
the pressure increase and
eliminate them before the next
measurement.
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
255).
If medically indicated, carry out
suitable treatment.
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
255).
If medically indicated, carry out
suitable treatment.
Change printer paper
98).
Change printer paper
98).
If medically indicated, carry out
suitable treatment.
Adapt alarm limits
255).
MEDUCORE Standard
(see
(see
(see
(see 10.3.7, p.
(see 10.3.1, p.
(see 10.3.1, p.
(see 4.7.3, p.
(see 4.7.3, p.
(see 10.3.1, p.
2
EN