18 Technical data
Specification
Storage/transport
(see "15 Storage",
page
354):
•
Temperature range
•
Humidity
•
Air pressure
•
Altitude above mean sea level
Unless stated otherwise, these
conditions also apply to all components
and all accessories.
Electrical connection (rated voltage)
Max. power consumption
Disconnection from line power
Operation with on-board power supply:
•
Rated voltage
•
Max. internal resistance of on-
board power supply
•
Maximum current consumption
Operating mode
Classification acc. to EN 60601-1:
•
Type of protection against elec.
shock
•
Degree of protection against elec.
shock (pulse oximetry sensor)
•
Degree of protection against elec.
shock (ECG electrodes)
•
Degree of protection against elec.
shock (defibrillation electrodes/
paddles)
•
Degree of protection against elec.
shock (NIBP cuff)
Degree of protection against
•
Ingress of solid objects
•
Ingress of dust
•
Ingress of water with harmful
effect
Mechanical resistance to falls
Electromagnetic compatibility (EMC)
acc. to EN 60601-1-2:
Radio interference suppression
Radio interference immunity
368
EN
MEDUCORE Standard
Device
-40 °C to +70 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
12 V to 15.1 V
30 W
Pulling out the power plug disconnects the device from line power
on all poles.
12 V
500 mΩ
< 3 A
Continuous duty
Protection class II
Degree of protection BF - defibrillation-proof
Degree of protection CF - defibrillation-proof
Degree of protection BF - defibrillation-proof
Degree of protection BF - defibrillation-proof
IP55: Protected against dust and water jets
1 m
Test parameters and limit values can be requested from the
manufacturer WEINMANN Emergency if required.
EN 55011
EN 61000-4 (parts 2 to 6, 8, and 11)
2