2 Safety
2.6.8 Printing
Warning
30
EN
MEDUCORE Standard
Risk of injury from falsified measured values during non-
invasive blood pressure measurement!
Cardiac rhythm disorders, arteriosclerosis, reduced perfusion,
diabetes, pregnancy, pre-eclampsia, arrhythmias (in newborns up
to 28 days), kidney problems, shaking, shivering or the use of a
cardiac pacemaker may impair the ability of the non-invasive
blood pressure measuring module to record correct measured
values. Evidence of the safety and efficacy of non-invasive blood
pressure measurement has not been obtained for patient groups
with these characteristics. This may injure the patient.
 Include the status of these patient groups when evaluating the
measured values of non-invasive blood pressure measurement.
 Only use non-invasive blood pressure measurement on patient
groups for whom evidence of the safety and efficacy of non-
invasive blood pressure measurement has been provided.
Risk of injury from overly frequent measurements!
Overly frequent measurements may lead to circulation problems
and patient injury.
 Select the measurement intervals to guarantee sufficient
perfusion.
 With extended NIBP measurements, check the position of the
NIBP cuff regularly and, if necessary, reposition it.
Risk of injury due to electric shock when the printer is
touched!
The printer's USB port is intended only for service purposes. A
connection between the printer and another device via the USB
port may lead to an electric shock if the printer is touched. This
may injure the patient or the user.
 Do not connect the printer to another device via the USB port.
 Only use the USB port for service purposes.
Risk of injury from electric shock when charging the printer
battery in the printer!
If the power supply unit and charger is defective, charging the
printer battery in the printer may lead to an electric shock if the
printer is touched. This may injure the patient or the user.
 Only charge the printer battery in the charging station for the
printer battery or the quadruple charging station for the printer
battery.