Weinmann MEDUCORE Standard2 Instructions For Use Manual page 135

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Shock not possible
The device carries out a cardiac rhythm analysis, charges for shock
delivery and outputs the message:
Voice prompt
Stand clear of the patient
Cardiac rhythm is being analyzed
If, based on the cardiac rhythm analysis, the device determines that
a shock is not required, the device outputs the message:
Voice prompt
Shock not recommended
1. Carry out CPR
Result
The patient does not have a cardiac rhythm which can be
defibrillated. The device guides you through CPR by means of voice
prompts, AED instructions, and the metronome. It warns you again
to stand clear of the patient after the preset time has elapsed
(120 s when device supplied), to allow it to carry out another
cardiac rhythm analysis.
Carrying out CPR
This section describes CPR in AED mode. When supplied, the
device carries out CPR with the following parameters; these,
however, can be adapted by the operator:
Setting
CPR phase
Duration
Pause for ventilation
AED instruction
Stand clear of the patient
Analysis
AED instruction
Shock not recommended
(see " Carrying out CPR", page
Patient group
Adult
120 s
5 s
MEDUCORE Standard
6 Operation
135).
Child
120 s
5 s
2
EN
135