Weinmann MEDUCORE Standard2 Instructions For Use Manual page 140

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

6 Operation
Result
6.7.2 Carrying out manual defibrillation using
Requirement
140
EN
MEDUCORE Standard
Failure of therapy due to defibrillation electrodes accidentally
coming loose during shock delivery!
Defibrillation electrodes accidentally coming loose during shock
delivery may lead to damage to the device and thus to the failure
of treatment. This may injure the patient.
 Ensure that the defibrillation electrodes are always connected
to the device during shock delivery.
9. Deliver a shock with the shock button
When doing so, please note:
If the shock button
discharges automatically after 30 seconds.
Only with the Print option: If the printer is connected to the
device and the Defibrillation report menu item is
activated
(see "13.9 Printer settings (only with Printing
option)", page
defibrillation report after shock delivery
Printing a defibrillation report", page
The patient has been given an electric shock.
paddles
This function is only available if manual mode has been enabled
and activated by the operator: Operator menu |
System settings | Enable options | Manual defibrillation
"13.12 System settings", page
If you are the operator of the device and have access to the
operator menu, you can disable manual mode: Operator menu |
System settings | Disable functions | Manual mode
"13.12 System settings", page
A charged battery is inserted in the device.
The device is switched on
page
118).
A patient group is selected
page
120).
2
is not pressed, the shock capacitor
332), the printer automatically prints a
342).
342).
(see "6.1 Switching on the device",
(see "6.4 Selecting patient group",
.
(see "6.16.4
201).
(see
(see

Advertisement

Table of Contents
loading

Table of Contents