8. Maintenance
8.1 Intervals
We recommend only having
servicing and maintenance
performed by Weinmann, the
manufacturer, or authorized
dealers.
36
Maintenance
EN
Both filters must be regularly checked for soiling.
•
The coarse filter should be cleaned once a week
and changed every 6 months.
•
The fine filter should be changed after 250 oper-
ating hours at the latest.
From reasons of hygiene we recommend:
•
replacing the pressure measurement tube
6 months and earlier when soiled;
•
replacing the complete mask system every 6 to 12
months depending on soiling;
•
replacing the exhalation system according to the
relevant operating instructions.
In addition, servicing must be carried out as a preventative
measure.
•
After every 5000 operating hours (maintenance
symbol
appears in display);
23
•
2 years at the latest (see servicing sticker on bot-
tom of unit).
Maintenance includes the following:
•
Filter change;
•
Cleaning unit;
•
Replacement of any defective parts;
•
Replacement of battery;
•
Complete check of unit functions and pressure
indicator.
after