Weinmann MEDUCORE Standard2 Instructions For Use Manual page 181

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Delayed therapy due to severely delayed reception of the
12-lead ECG!
If the infrastructure of the monitoring station/hospital is not
configured by the operator to receive 12-lead ECGs, receipt of the
12-lead ECG may be severely delayed. This may lead to delayed
therapy and injure the patient.
 Configure e-mail delivery 12-lead ECG correctly and check it
(see "12.5 Configuring e-mail delivery 12-lead ECG (only with
E-mail delivery 12-lead ECG option)", page
 If transmission time is unexpectedly high: Contact the system
administrator of the e-mail recipient.
14. If desired (only with E-mail delivery 12-lead ECG option): Press
the Send function button.
15. Select the recipient's name and send the 12-lead ECG to a
predefined recipient by e-mail
delivery 12-lead ECG (only with E-mail delivery 12-lead ECG
option)", page
Following successful delivery, the symbol
display. If delivery is unsuccessful, the symbol
the display and an alarm is displayed
messages", page
(see "12.5 Configuring e-mail
299).
271).
MEDUCORE Standard
6 Operation
299).
appears in the
appears in
(see "11.2 Alarm
2
EN
181