Weinmann MEDUCORE Standard2 Instructions For Use Manual page 331

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

13.8.2 Possible values
Parameter
Adult
Initial cuff
Child
pressure
Infant
Interval
-
Venous stasis
-
pressure
Venous stasis
-
duration
Possible
Description
values
120 mmHg -
280 mmHg,
in increments of 5 Here you can set the NIBP
80 mmHg -
cuff pressure to which the
170 mmHg,
device must inflate the NIBP
in increments of 5
cuff for a particular patient
group.
60 mmHg -
140 mmHg,
in increments of 5
Off
0:30 min
0:45 min
1:00 min
1:15 min
1:30 min
Here you can set after how
1:45 min
long another NIBP
2:00 min
measurement is to be started
3:00 min
automatically.
5:00 min
10:00 min
15:00 min
30:00 min
60:00 min
Here you can set the pressure
60 mmHg -
to which the NIBP cuff is
140 mmHg,
inflated for venous stasis
in increments of 5
(IV function button).
Here you can set for venous
stasis (IV function button)
10 s - 120 s
the time for which pressure is
to be maintained in the NIBP
cuff.
MEDUCORE Standard
13 Operator menu
Factory
setting
180 mmHg
120 mmHg
90 mmHg
Off
80 mmHg
60 s
2
EN
331