Recommended separation distances between portable and mobile HF
telecommunications devices and MEDUCORE Standard
2
MEDUCORE Standard
variables are controlled. The customer or user of the MEDUCORE Standard
interference by maintaining a minimum distance between portable and mobile HF telecommunications
equipment (transmitters) and MEDUCORE Standard
telecommunications equipment, see below).
Separation distance depending on transmission frequency in m
Rated power of
150 kHz -
HF device in W
80 MHz
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
Recommended separation distances between portable and mobile HF
telecommunications devices and MEDUCORE Standard
Separation distance depending on transmission frequency in m
When used as a defibrillator
Rated power of
HF device in W
150 kHz -
80 MHz
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
Risk of injury from overly powerful high-frequency
interference signals or electromagnetic fields!
Overly strong high-frequency interference signals from portable
high-frequency communications equipment (e.g. radios) or
electromagnetic fields may lead to incorrect analysis, incorrect
measurement results, and incorrect alarms, and thus impair the
functioning of the device and injure the patient.
 Maintain separation distances.
 With portable high-frequency communications devices,
maintain a minimum distance of 30 cm from the device,
components, and accessories.
is intended for use in an electromagnetic environment in which HF interference
150 kHz -
80 MHz in the
ISM bands
0.12
0.38
1.2
3.8
12
150 kHz -
80 MHz in the
ISM bands
0.27
0.66
2.7
6.6
27
2
2
(depending on the output power of the HF
When used as a monitor
80 MHz -
800 MHz
0.4
1.3
4
13
40
2
Defibrillator: No unintended
energy delivery
80 MHz -
800 MHz
0.06
0.15
0.6
1.5
6
MEDUCORE Standard
18 Technical data
2
can avoid electromagnetic
800 MHz -
2.5 GHz
0.77
2.4
7.7
24
77
800 MHz -
2.5 GHz
0.12
0.38
1.2
3.8
12
2