6 Operation
154
EN
MEDUCORE Standard
•
The triangles
13. Position paddles on the upper torso in accordance with
currently applicable guidelines.
14. Briefly press and release the charge button
the APEX paddle.
When doing so, please note: The shock capacitor can only be
charged if the user menu is not activated.
The charge progress bar appears. A rising charging tone
sounds until the device is ready to deliver the shock.
15. To cancel shock charging: Cancel shock charging by pressing
the Cancel function button or by switching to another mode.
16. Check the ECG leads to see whether cardioversion is still
indicated.
2
must mark every R wave reliably.
(CHARGE) on