Advertisement

New risks from integrating the device in an IT network!
When integrating the device in an IT network, new risks may arise
for patients, users or third parties which need to be determined,
analyzed, and managed.
 As the operator of the IT network, reassess changes to the IT
network. When doing so, please note:
Modified IT network configuration
Connection of additional elements to the IT network
Removal of additional elements from the IT network
Updates of devices connected to the IT network
Upgrades of devices connected to the IT network
Delayed therapy due to integrating the device in a network
with a login page!
If the device is integrated in a network with a login page, it is not
possible to upload a 12-lead ECG by e-mail.
 Do not use a network with a login page.
2. Select the desired network.
When doing so, please note: The device must only be
connected to an encrypted network. A password always has to
be entered on an encrypted network.
3. Confirm with the OK function button.
4. Enter the password for the desired network.
The device connects to the desired network.
5. In the event of problems setting up the network: Contact
system administrator.
Result
A network has been configured for the device automatically and
specified as the default network.
From now on, the device always automatically sets up a connection
to the default network set up when a WiFi connection is required.
Following successful data transmission, the WiFi connection is
disconnected.
12 Configuration by the operator
MEDUCORE Standard
2
EN
295