8 Hygienic reprocessing
Manual
cleaning
(only
Part
necessary in
the event of
visible dirt)
Bags and portable unit
Printer case
Wipe down
smooth surfaces
with neodisher
MediClean forte
(Dr. Weigert).
Dose:
MCS2-Bag
10 ml/l
protective
transport bag
Wipe down all
surfaces at least
2x until they
look clean.
Wipe down with
neodisher
MediClean forte
(Dr. Weigert).
Dose:
Retaining plate
10 ml/l
Wipe down all
surfaces at least
2x until they
look clean.
Portable unit
Follow the instructions for use of the portable unit
246
EN
MEDUCORE Standard
Disinfection
by wiping
®
Not permitted
®
Wipe down all
surfaces with
Incidin™
OxyWipe S
(Ecolab).
Exposure time:
5 min
2
Disinfection
Mechanical
by
reprocessing
immersion
Wash in a
washing
machine with no
spin cycle.
Prewash: 5 min
at 20 °C
Main wash:
Not permitted
30 min at 30 °C
Cleaning agent:
Turbo Usona
(Ecolab)
Disinfectant:
Turbo
Oxysan (Ecolab)
Not permitted
Not permitted
Sterilization
Not permitted
Not permitted