11 Alarms and faults
Alarm
SD card defective
SD card full
SpO
sensor defective
2
signal quality *
SpO
2
Update operator menu access
code
(Alarm deactivated after 10 s)

11.3 Faults

278
EN
MEDUCORE Standard
Cause
SD card write/read error
SD card full
Pulse oximetry sensor connecting
cable defective
Pulse oximetry sensor defective
SpO
signal quality < 40 %
2
Access code for the operator menu
is 000000 (status when device
supplied)
* If the device outputs this alarm, it will no longer output a
simultaneously occurring SpO
quality drops to < 20 %, the device will also stop displaying the
SpO
measured value and the plethysmogram.
2
If you are unable to remedy faults with the aid of the table, you
should contact the manufacturer WEINMANN Emergency, or
technicians expressly authorized by WEINMANN Emergency, to
have the device repaired. To avoid more serious damage, do not
continue using the device.
2
Remedy
Insert new SD card
92), switch off device
119), and switch device back on
after at least 30 s
Resume session and do not record
any more data.
Insert new SD card
92), switch off device
119), and switch device back on
after at least 30 s
Erase data on current SD card.
Replace pulse oximetry sensor
connecting cable.
Replace pulse oximetry sensor.
Attach pulse oximetry sensor to
the patient correctly
157).
If medically indicated, carry out
suitable treatment.
Wait 10 seconds.
Change operator menu access
code
(see "13.12 System
settings", page
measured value alarm. If signal
2
(see 4.6.1, p.
(see 6.2, p.
(see 6.1, p.
118).
(see 4.6.1, p.
(see 6.2, p.
(see 6.1, p.
118).
(see 6.9.1, p.
342).