11 Alarms and faults
11.3.7 Printing
Fault
Printer not connecting to the
device
Connection between device and
printer fails
Connection between printer and
device is extremely slow
Special characters in printout or
printer fails in airplane mode
286
EN
MEDUCORE Standard
Cause
E-mail delivery 12-lead ECG or
network data transmission
occurring in parallel
Poor reception quality
Feed button (FEED) pressed for
longer than 10 s
Poor reception quality
Electromagnetic radiation in
airplane mode
Printer is at the limit of device
range
2
Remedy
Wait until e-mail delivery 12-lead
ECG or network data transmission
is finished. If necessary: Switch the
printer on again.
Put printer within range of the
device.
Do not press feed button (FEED)
for longer than 10 s.
Put printer within range of the
device.
Do not use the printer in airplane
mode.
Reduce the distance between
device and printer.