11.3.6 Non-invasive blood pressure measurement
Fault
Implausible measured values
NIBP measurement cannot be
started
NIBP measurement keeps failing
NIBP cuff cannot be fully inflated
(NIBP monitoring)
Cause
NIBP cuff attached incorrectly
NIBP cuff leaking
Adapter tube for connecting
disposable NIBP cuffs not
connected correctly
Adapter tube for connecting
disposable NIBP cuffs leaking
Unsuitable patient position
NIBP module defective
NIBP cuff size not suitable (too
large or too small)
Device defective
NIBP cuff and/or NIBP connecting
tube not connected correctly
NIBP cuff is constricted and cannot
be inflated
Movement results in artifacts
NIBP cuff or NIBP connecting tube
damaged
Connection of NIBP cuff and/or
NIBP connecting tube interrupted
Adapter tube for connecting
disposable NIBP cuffs not
connected correctly
Device defective
11 Alarms and faults
Remedy
Reattach NIBP cuff
182).
Replace the NIBP cuff.
Connect adapter tube for
connecting disposable NIBP cuffs
correctly.
Replace the adapter tube for
connecting disposable NIBP cuffs.
Reposition the patient.
Have the device repaired.
Use a suitable NIBP cuff size.
Have the device repaired.
Connect NIBP cuff and/or NIBP
connecting tube correctly.
Remedy NIBP cuff constriction.
Reduce movements.
Replace NIBP cuff and/or NIBP
connecting tube.
Connect NIBP cuff and/or NIBP
connecting tube.
Connect adapter tube for
connecting disposable NIBP cuffs
correctly.
Have the device repaired.