6 Operation
Result

6.10 6-lead ECG monitoring

6.10.1 Preparing 6-lead ECG monitoring
Requirement
164
EN
MEDUCORE Standard
5. If necessary: Make the following SpO
menu
(see "10.3.4 SpO
Adapt the feed rate of the plethysmogram.
Give audio pulse tone output priority over heart rate tone
output so that the tone level is dependent on oxygen
saturation even when the ECG cable or defibrillation
electrodes are connected to the patient.
6. If necessary: Set alarm limits in the user menu
Alarm settings", page
Alternatively:
Set automatic alarm limits in the application menu
Application menu", page
7. If necessary: Deactivate the pulse tone with the heart rate tone/
pulse tone function button
8. If there are artifacts in the SpO
(bar in the SpO
oximetry sensor on the patient's limb.
Pulse oximetry monitoring is carried out.
The device is switched on
page
118).
A patient group is selected
page
120).
Risk of injury due to live connection for 12-lead ECG extension
cable!
The connection for the 12-lead ECG extension cable may be live
and lead to an electric shock if contacted. This may injure the user.
 Always seal the 12-lead ECG extension cable connection with
the protective cap if no 12-lead ECG extension cable is
connected.
2
settings", page
2
255).
248).
.
curve or if signal quality is poor
2
parameter field): Reposition the pulse
2
(see "6.1 Switching on the device",
(see "6.4 Selecting patient group",
settings in the user
2
261):
(see "10.3.1
(see "9