Uploading Function Check To Weinmann Connect - Weinmann MEDUCORE Standard2 Instructions For Use Manual

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

5 Function check
Result
5.3 Uploading function check to
Requirement
112
EN
MEDUCORE Standard
Display
Device not ready
for use
Device ready for
use
The service symbol
flashes in the start
menu.
23. Finish function check with the OK function button.
The start menu appears.
24. Connect the ECG cable to the ECG connection for ECG cable.
25. Connect the NIBP cuff to the NIBP connection for the NIBP
connecting tube using the NIBP connecting tube.
The function check is complete. The device is ready for use.
WEINMANN Connect
A default network has been configured
default network", page
The device has been registered with WEINMANN Connect
"12.4 Registering the device with WEINMANN Connect",
page
298).
A function check has been carried out.
The status report of the function check is displayed.
2
Meaning
Function check failed
or
Function check canceled
Function check passed,
but note about service
due
292).
Action
Repeat the function
check.
Contact the
manufacturer or a
technician expressly
authorized by
WEINMANN Emergency.
Use device without
restriction, but have
device checked.
(see "12.3 Configuring
(see

Advertisement

Table of Contents
loading

Table of Contents