3 Description
Status indicator on
the battery
58
EN
MEDUCORE Standard
Status indicator on
the device display
2
Meaning
Battery status approx. 10 % - 40 %
Battery status < 10 %
In the display:
•
The last remaining segment in the battery status
symbol is red.
•
Battery weak appears in the display.
•
In AED mode, the device outputs: Battery weak.
Battery is deeply discharged. Charge battery in the
device for 24 h. After 24 hours:
•
Green LED is on: Battery fully charged and ready
for use
•
Red LED or no LED on: Battery defective. Replace
battery.
Battery is empty
Battery empty appears in the display and in AED
mode, the device outputs:
Battery empty.
The device can be used for about another 15 minutes.
Battery is defective. Replace battery.
•
Battery is defective.
or
•
No battery.
or
•
Battery not at suitable temperature.
Green arrow: Battery charging