3 Description
Status indicator on
the battery
58
EN
MEDUCORE Standard
Status indicator on
the device display
2
Meaning
Battery status approx. 10 % - 40 %
Battery status < 10 %
In the display:
The last remaining segment in the battery status
symbol is red.
Battery weak appears in the display.
In AED mode, the device outputs: Battery weak.
Battery is deeply discharged. Charge battery in the
device for 24 h. After 24 hours:
Green LED is on: Battery fully charged and ready
for use
Red LED or no LED on: Battery defective. Replace
battery.
Battery is empty
Battery empty appears in the display and in AED
mode, the device outputs:
Battery empty.
The device can be used for about another 15 minutes.
Battery is defective. Replace battery.
Battery is defective.
or
No battery.
or
Battery not at suitable temperature.
Green arrow: Battery charging