Weinmann MEDUCORE Standard2 Instructions For Use Manual page 313

Monitor/defibrillator
Hide thumbs Also See for MEDUCORE Standard2:
Table of Contents

Advertisement

Parameter
Adult AED settings/Child AED settings
CPR voice
prompts
Metronome
Audio outputs
CV ratio
Metronome
frequency
Possible
Description
values
Here you can set whether the
Activated
voice prompts for guiding
chest compressions and for
ventilation are to be given.
Deactivated
The voice prompt Ventilate
twice is only given if the
metronome is set to the
rhythm 15:2/30:2.
off
15:2/30:2
Here you can set the
metronome rhythm.
Here you can set the
compression/ventilation
ratio.
30:2
15:2
100/min to
Here you can set the
120/min
metronome frequency.
MEDUCORE Standard
13 Operator menu
15:2: After
15 metronome beats,
there is a pause to allow
for ventilation.
30:2: After
30 metronome beats
there is a pause to allow
for ventilation.
Factory
setting
Deactivated
15:2/30:2
AED mode
adult: 30:2
AED mode
child: 15:2
100/min
2
EN
313